        INSTRUCTIONS

        FA/JA  ELECTRONIC BALANCE

        OPERATION

         

        CONTENT

             1. Outline                                     (1)

             2. Main technical Specifications                   (2)

             3. Drawings for Balance Installation                (4)

             4. Operating                                   (8)

              A. Preparation                                 (8)

              B. Operation                                   (8)

              C. Start                                       (8)

              D. RNG Weighing                              (9)

              E. UNT Unit Change                            (9)

             F. INT Integrating Time Adjustment                (9)

              G. ASD Sensitivity Adjustment                    (10)

              H. Balance Calibration                          (10)

              I. PRT Output Mode Setting                      (10)

              J. COU Count Function                          (11)

              K.Weighing,Taring,Add Objects,Read Devitions      (11)

             5. MaintenanceandTroubleshooting               (12)

             6. Dats InterfaCe                              (14)

             7. Serial Single ( RS232C )                    (14)

             8. Random Accessories                             (15)

        can be connected to the microprocessor and various printers.No.Standard Parallel Output.The pin and the corresponding signanls are as follows:

        PIN      SIGNAL      ILLUSTATE 
           1        SI           input signal

        2       GND

        3        SO          serial output signal(baud rate is 1200)

        4       GND

        7.Serial Single (RS232C)

        The connection method between the scale and computer serial port is as follows:

        Computer(9pins)                   Scale(9pins)

        2                              3

        3                              2

        5                              5

        1) The baud rate of  the serial port of scale is 1200.

        2) The date format is 10digits, one start digit is(0) ,8 digits are date positions (ASCII Code,low digits in the front),1 stop digit.

        3) No odd-even check 
        4) The date will be output continuously without any special reading commands A detailed output frame is as follows:

        1

        2

        3

        4

        5

        6

        7

        8

        9

        10

        11

        12

        13

        14

        15

        16

        Type

        Space

        */Space

        +/-

        Date

        Date

        Date

        Dot

        Date

        Date

        Date

        Date

        Unit1

        Unit2

        CR

        LF

        8  Random Accessories

        1.  200g(and100g)calibration weights                1 piece

        2.  Wire with both-side plug                        1 piece

        4 . Weights tweezers                             1 piece

        5.  Handkerchief                                 1 piece

        6.  Operation manual                              1 piece

        7.  Certificate                                   1 piece

         

        SN

        TROUBLE

        CAUSE

        REMEDY

        6

        Display remains at a certain digit or indicates nonsense symbol

        ·Instant interference 

        ·Wreng line voltage

        ·Turn on the balance once again or replug the power cord 

        ·Chang to normal line voltage

        7

        The stable mark 0on  the left of the  display doesngo out    

        ·A higher balance sensitivity 

        ·undesirable environment such as strong air flow or vibration

        ·Set to lower sensitlivity 

        ·Improve the environment

        8

        Remain at Waiting Status————

        ·Thr balance position is nocorrect, e.g. With strong air flow, vibratioor great fluctuation of room  temp. 

        ·The selected sensitivity is toohigh

        Improve the environment  ASD-3

        9

        Cal Err 

        Displays

        ·There is an object on the pan befpre calibration

        ·not clear before calibration

        ·Depress CAL before the

        ·Take away the object,clear and recalibrate 
        ·Clear and recalibrate

        10

        Err-1  Err-2

        Display

        ·Instant interference

        ·Something wrong with thebalance

        ·Ture on the balance once again

        ·Send it to the service center

        11

        The weighing unit on the right dosent display

        ·not calibrated

        ·The calibrated number in the internal memory of the balance has been erased

        ·Calibrate the balance

        12

        Cou-Err  Display

        ·No Constant be set before operating counting function

        ·Overload when setting constant

        ·Underload when setting constant

        ·Set average number before operating counting function

        6.Date Interface

        Date interface adopts the standard 9-core RS-232C sDcket. The date interface is provided with RS232C universal two - way serial interface, which

        Thank you for busying the type FA/JA series electronic balance frorn our company. For correct and safe fixing and operation, as well as full function apply, we suggest that you should read tha users’ manual before using it Thanks 
         The balance is multifunctional electronic balance using MCS-5l series SCM This balance has gram, cant, ounce for users to select (cant and ounce for export) besides having automatic calibration, integral time adjustment and stability adjustment functions This balance has RS232C data serial duplexing port to connect with microcomputer and serial port punter This balance is used for fast exact weight measuring widely in enterprise, university, graduate school lab as an ideal lab instrument.
         Finally, after reading the manual, please keep it carefully for referring.

         

        2  Main technical Specifications

               

        Model

        FA1004

        FA1104

        FA1604

        FA2004

        FA2104

        Accuracy Degree

        Weighing Range(g)

        100

        110

        160

        200

        210

        ReadlngAccuracy(mg)

        0.1

        0.1

        0.1

        0.1

        0.1

        Taring Range(g)

        0~100

        0~110

        0~160

        0~200

        0~210

        Repeatability Standard Deviation(g)

        ±0.0002

        ±0.0002

        ±0.0002

        ±0.0002

        ±0.0002

        Linear Error(g)

        ±0.0005

        ±0.0005

        ±0.0005

        ±0.0005

        ±0.0005

        Stable  Time (Typical)(s)

        6

        6

        8

        8

        8

        Integrating  Time (Adjustable)(s)

        2/4/8

        2/4/8

        2.5/5/10

        2.5/5/10

        2.5/5/10

        Pan Dia.mm)

        Φ80

        Overall Dimensions (mm)

        360×217×345

        Net Weightkg)

        6.8

        Power Supply

                  220V

        +22V

        -33 V

        50Hz110V ±11V  60Hz

        Power Consumption

        (V .A)

        15

        Auto-Cal Weight

        Range(g)

        100

        100

        100

        200

        200

        Warm-upTimemis)

        180

        180

        180

        180

        180

        weight the negative value of the weight dispIay.Put the object to be weighed on the pan.The corre sponding plus or minus deviution will display, comparing the weighed object with the reference weight

        UNLOAD

        Loose the screwn of the bottom cover,raveal the hook.Put the balance on a woking table with a hole. Level and calibrate the balance .An object can be weighed with the hook.

        5.Maintenance and troubleshooting

        MATNTENANCE OF THE BALANCE

        The batance should be used carefully.Clean the paand the case frequently with soft cloth and toothpaste.Don,t wipe the batance with strong agent.

        TROUBLESHOOTING 

        SN

        TROUBLE

        CAUSE

        REMEDY

        1

        No ligbt on Display

        ·Power is not on

        ·Disptsy switch is not on 

        ·Instant interrference  

        ·Check and turn it on again 

        ·Depress ON button 

        ·Switch on again and replug power 

        2

        Displsy the upper hstf only

        ·Overload

        ·The calibration in the

        intemal memory may ba damaged

        ·The pan is not installed correctly    

        ·Reduce the load immediatel

        ·Recalibrate according to the above procedures. About 8 seconds after the standard weight has been put on,the calibrated result may be displayed. A certain stable time is necessary.

        ·Take out the pan and reinstall it.

        3

        Display the lower half  Only———

        ·without pan,too light 

        ·he pun is not  installed correctly

        ·Reinstall the pan

        4

        The weighing result is not stable (date changed swiftly)

        ·due to air flow 
        ·the working table it net stable 
        ·the integrating time is too short 
        ·Roam tetnpeealtsre flucttsates

        ·Cheek the windproof cover tu see if it is closed. 

        ·Change to a longer integrating time

        5

        The retult is not correct

        ·not zeroing before weighing 

        ·Use the balance without calibration or the cal weight is not accurate 

        ·The line vollage is too low or net correct

        ·Depress TAR butto
        ·Reclibrate 
        ·Change to the correct voltage

         

        if a normal weighing function is required ,release when COU-00 display, and waiting status“……” will be indicated ,at last0.000g weighing status will be indicated .lf the average value of five is required, release when COU-5 displays, then put5 objectson  the pan Depress CAL once again. “……”waiting status displays ,and about several seconds later,5 will be displayed. Take away the objects to be weighed ,zero displays. At that time the counting of the same objects to be weighed can be done (Attention :the weight of the objects to be weighed must not be greater than the maximum weighing rang of the balance). If you average over 10,25 even SO objects ,then the accuracy of counting will be higher Depress TAR,“0”will be indicated in the display and counting can be done at that time.

        K . Weighing ,Taring ,Add Objects ,Read Devitions

        WEIGHING

        After the selection of the above modes(all the modes can be used for weighing after the power is off ,due to the memorv function of the balance ),depress TAR,zero will display.Put the object to be

        weight on the pan.When the number is stable, i.e.“0”on the left of the displaygoes out,the number 

        display goes out,the number displayed will be the weight of the object.

        TARING

        Put the container on the pan, the weight will display.Depress TAR,zero displays. That is taring. Put the objects to be weighed into the container, the value displayed is the net weight of the object to be weighed

        ACCUMULATIVE WETGHTNG

        Pat the objects be weighed on the pan one by one with taring method and tare and clear for each one.Take away all the objects to be weighed , the absolute value display in the total weight of the objects to be weighed.

        ADD OBJECTS

        Set the mode of INT-0 and put the container on the pan , then tare. Add the objects to be weighed (liquid or loose objects) into the container one by one the continuous reading value can be obtained quickly. when the added objects reach the required“0”on the left of the display goes out and the number display is the weighing value required by the user. When adding the mixed objects.the net weight of each object can be measured by taring method.

        PEAD DEVIATIONS

        Put the reference weight(or rample) on the pan and tare . then take off the reference

        Model

        JA1003

        JA1203

        JA2003

        JA3003

        JA5003

        Accuracy Degree

        Weighing Range(g)

        100

        120

        200

        300

        500

        ReadlngAccuracy

        (mg)

        1

        1

        1

        1

        1

        Taring Range(g)

        0~100

        0~120

        0~200

        0~300

        0~500

        Repeatability Standard Deviation(g)

        ±0.001

        ±0.001

        ±0.001

        ±0.001

        ±0.001

        Linear Error(g)

        ±0.002

        ±0.002

        ±0.002

        ±0.002

        ±0.002

        Stable  Time (Typical)(s)

        6

        6

        6

        6

        6

        Integrating  Time (Adjustable)(s)

        2/4/8

        2/4/8

        2/4/8

        2/4/8

        2/4/8

        Pan Dia.mm)

        Φ80

        Φ125

        Overall Dimensions (mm)

        360×217×345

        Net Weightkg)

        6.8

        Power Supply

        220V

        +22V

        -33 V

        50Hz 、110V ±11V  60Hz

        Power Consumption

        (V .A)

        15

        Auto-Cal Weight

        Range(g)

        100

        100

        200

        200

        200

        Warm-upTime

        mis)

        60

        60

        60

        60

        180

        Note:The balance is turned on when the power is on. The panel switches activate display only. The power plug should be removed if the balance is not used for a long time.If it is used daily, the power needn’ t be turned off.The only thing needs to be done is to turn off the display.It is not necessaly to warm up due to the power on,so it can be used at any time.(“Aong time”means more than 5 days.)

         

        3.Drawings for Balance Installation

        3        

        2                            1.panel 

                                  2.gradienter 

        3.scalepan

        1

                                                              

                                                                     

                      

           

                         

        1                            1. Data Interfale       

                                    2. Transformer        

              3.Electrical Outlet

            3

         2

                                            

        Take away all the objects to be weighed on the pan. Set RNG-60,INT-3, ASD-2, and unt-g. Depress slightly TAR for clear thebalance.

        BALANCE CALIBRATION

        Depress slightly CAL.CAI-200 will be on the flash-display. Therefore, put the prepared 200g standard weight on the pan ,the flash-display will stop.

        After several seconds, 200.000g will be displayed. Remove the calibration weight,the display should indicate 0.000g. if not clear once again and repeat the above procedures. (Attention:More than two times of calibration are mcommended in order to get accurate result.)

        The display sequence of the calibration is shown as the figure.

                

          

        I . PRT Output Modde setting

        Deprss PRT and hold there will be four modes displayed circularly for user to select at will.. 

        PRT-O is the mode of indefinite time output ; slghtly depress PRT once ,the weighing 

        result will be.out on the output interface once.

        Attention: At this time you should depress the button slisghtly and quickly, otherwise, the next output mode will be displayed

        PRT-1 output once every half a minute

        PRT-2 output once every minute

        PRT-3 output once every two minutes

        The setting of PRT is the same as RNG

        J. COU Count Function

        The balance has a function of counting. The average number are 5,l0,2nd 50,total of four .The Setting of Range of Average Numbers

        Only depress COU and hold ,the display will be circulating continuously as the following.

        It means the average value of (5,10,25,50)objects respectively

         

        E.UNT Unit Change

        If the accuracy of the reading needs to be 0.1mg release when the display is mg-60; and the appears waiting state...,the weighing state will be shown at last.

        Depress ERY button and hold until the display is shown as the following figure circulating continuously.

        “g” means the value is expressed in gram .“—” means the value is expressed in metric carat

        “y” means the value is expressed in oz.t.The unit will be set same as RNG.   

        F. INT Integrating Time Adjustment

        Four modes of integrating time can be selected circularly as shown in the figure    

        The corresponding integrating time is as follows:

        INT-0 fast, NT-1 shorter, INT-2 short, INT-3 longer

        The selection of the integrating time will be the same as RNG

        G. ASD Sensitivity Adjustment

        Same as INT interrating time adjustment button: there are four circulating modes

        provided for sensitivity adjustment.

        ASD0  the highest   ASD1  high   ASD2  higher  ASD3  low

        ASD-0 is used for test run and not to be usd by the users.

        The selection of sensitivity is the same as RNG. 

         Here is a list of ASD usesd with INT for users reference only.  

        The fastest weighing: INT-1 ASD-3   

        Normal: INT-3 ASD-2

        With undesirable environment INT-3 ASD-3  

        H. BALANCE CALIBRATION

        Normally, the balance should be calibrated after a long period of storage, movement, change of environment or in order to obtain precise measurement.

        Preparation for the balancce calibration:

        Single Measuring Range

            

                1

        2

        1.Cover Honrd         2..Pothook

         

        4. Operating

        A. Preparation

        ·Unpack the box and remove all packing. Take out buffer sponge in the wind proof cover and install pans 

        ·Put the balance on a stable working table free form vibration, sunshine and air flow.

        ·Ambient temperature:

        20℃±5 for the first class balance with a fluctuation of temperature not greater than 1℃/h;

        20℃±7.5 for the second class balance with a fluctuation of temperature not greater than 5℃/h;

        ·Relative humidity:50%-75% for the first class lance, 50%-80% for the second class balance.  

        +22V

        -33 V

        ·Working voltage:  220V 50Hz 、110V ±11V  60Hz

        B. Operation

        ·Check the level meter before opertion,lf the bubble is not in the center, adjust the level legs to make the bubble in the center..

        ·The balance adopts soft touch button, so it can be controlled with multikey board.lt is easy to operate .Function change and selection can be realized simply by depressing the corresponding buttons.

        C. Start

        ·Select an appropriate line voltage and set the voltage switch to the corresponding position.

        ·Turn the power on and the balance is ready to run(but the display doesnt work ),normally one hour after power is on, the display can be operated.

        ±

        。

        8888888

        FA2004

        %

        g

        Function of  the keyboard:

        ONDISPLAY ON

        Depress slightly the button ON the display will be lighted..

        Check the function of the display ,about 2 seconds later ,the model of the balance will be displayed.

        For example:

        then the weighing mode. 

        0.0000g or 0.000g

                                             or

        OFF DISPLAY OFF

        Depress slightly the buttonOFF ,the display will go out. If the balance will not be used for a long time ,the power plug should be taken off. 

        “TAR” Clear and Tare  

        Put the container on the pan,the weight of the container should be displayed:

        Depress “TAR” button,the display will go out and the display will be all zeroes. Taring ia completed:   

        When the container is removed, a negative value of the container weight will be displayed.

        Depress“TAR”again, the display will be all zeroes,i.e.the balance clears.

        D.“RNG”Weighing

        The balance has two reading accuracles: 0.1mg within range 0-60g; if the range is above 60g the balance changes its accuracy to 1mg automatically. But through the function of taring of 0-210g,the accuracy will be 0.1mg for reading step if the total

        weight is below 210g,(the analysis quantity is below 60g);i.e.if the weight of the contaner is above 60g, depress slightly “TAR”,taring the weight of the container first,then weigh the object≤60g), the reading accuracy is still 0.1mg.

        set the Weighing Range

        Depresa“RNG”and hold until the display is shown as the following figure. circulating continuously.  